Ion Milling – ORION-8-IM System
Ion milling is a dry etching method used for creating thin film patterns. The process of ion milling takes place in a vacuum chamber in which ions Ar are accelerated by the ion gun towards the sample and bombard the areas of the sample uncovered by a mask. The collisions of the Ar ions with the surface of the film gradually remove surface atoms of the surface. This method is superior to wet chemical etching and lift-off process because it does not involve chemical reactions or film straining. It allows to remove of layers of materials with atomic precision and to form nano-sized patterns.
SPECIFICATIONS AND CAPABILITIES OF ORION-8-IM
- The sample stage is water-cooled and rotatable.
- It can be tilted out of horizontal position up to vertical position.
- The sample stage can accommodate samples with linear size up to 6”.
- Samples are loaded manually through the front door.
- The maximum RF power of the ion gun is 600 W. The tuning of the gun power is manual.
- The maximum etching rates depend on the material and are typically in the range from 0.1 to 1 nm/s.
- All kinds of materials can be etched.
- Samples containing toxic materials are prohibited.
- Table of the etching rates for different materials is provided.
- Different non-toxic gases e.g. O2, Ar, N2 can be used for ion milling (Ar is the most common).
- The pumping system consists of a dry diaphragm backup pump and the turbomolecular pump with the pumping speed of 700 l/s (Pfeiffer Vacuum)
Sample Data
User Rates
Rates will be based on an hourly or per sample basis and will be decided after consultation with the customer.
Hourly rates for instrument use
Assisted Use: $15/hr.
Training including cleanroom training:$50/hr.
Self-Use: $10/hr.
Consumables charged at cost.
Contact Us
Dr. Leszek Malkinski
Email: lmalkins@uno.edu